SDXNS200H Combustion Ion Chromatography System (CIC)

The SDXNS200H Combustion Ion Chromatography System (CIC) is an integrated analytical platform for the qualitative and quantitative determination of fluorine, chlorine, bromine, iodine and sulfur in solid and liquid samples. It combines programmable high-temperature pyrolysis, steam-assisted conversion, absorption and collection, automatic volume adjustment and dilution, and ion chromatographic separation in a fully coordinated workflow.

Designed for samples including coal, biomass, petrochemicals, paper products, degradable materials, electronic and electrical products, soil, ores and cement, the system provides a measuring range of 5 ppm to 1% in the standard configuration and 0.2 ppm to 1% in the ultra-low-level configuration. Automated sample handling, flexible combustion atmospheres, simultaneous multi-element analysis and integrated data management make the SDXNS200H suitable for laboratories requiring efficient, sensitive and reliable halogen and sulfur analysis.

PFAS Application

SDXNS200H can be used as a rapid, non-targeted fluorine screening tool for materials suspected of containing per- and polyfluoroalkyl substances. Its software includes configurable pyrolysis and steam-introduction parameters for perfluorinated compounds, enabling method optimization for different sample matrices. By determining the overall fluorine content released during combustion, the SDXNS200H helps laboratories screen samples, identify potentially fluorine-rich materials and prioritize samples for further compound-specific analysis.

Key Features

Fully Automated CIC Workflow

Once the test parameters are configured, the system automatically performs sample introduction, furnace-chamber closing, staged pyrolysis, analyte collection, volume adjustment, dilution, IC injection, chromatographic analysis and result calculation.

Multi-Element Analysis

The system determines F, Cl, Br, I and S, while fluorine, chlorine, bromine and sulfur can be analyzed simultaneously during continuous testing.

High Sensitivity and Wide Measuring Range

The standard version covers 5 ppm to 1%, while the ultra-low-level version extends the lower limit to 0.2 ppm, supporting both routine and trace-level analysis.

Broad Sample Compatibility

Selectable argon, oxygen and steam-assisted atmosphere conditions, together with support for solid and liquid sample introduction, allow the system to handle a wide variety of matrices.

Efficient Sample Throughput

The adjustable pyrolysis time is no more than 15 minutes. During continuous operation, the average analysis time is also no more than 15 minutes per sample.

Reliable Ion Chromatographic Separation

Ion chromatography separates the target ions in the absorption solution, minimizing mutual interference and providing high sensitivity, good repeatability and low detection limits.

Complete Software and Data Management

The workstation integrates instrument control, chromatogram acquisition, peak processing, calibration, result calculation, historical-data management, report generation, Excel export, user authorization and database backup. It can also interface with a LIMS.

Maintenance-Friendly Structure

The magnetic-coupling sample introduction mechanism can be removed without tools. The open-type furnace simplifies combustion-tube inspection, while the modular IC section supports both automated injection and direct liquid-sample analysis.

Applicable International Standards and Methods
StandardStandard TitleApplicable Samples
ASTM D7359Standard Test Method for Total Fluorine, Chlorine and Sulfur in Aromatic Hydrocarbons and Their Mixtures by Oxidative Pyrohydrolytic Combustion Followed by Ion Chromatography Detection (Combustion Ion Chromatography—CIC)Aromatic hydrocarbons and their mixtures
ASTM D8247-19Standard Test Method for Total Fluorine in Coal and Coke by Pyrohydrolytic Extraction and lon Selective Electrode or lon Chromatograph Methods (Withdrawn 2020)Petroleum Products and Lubricants
DIN 51723Determining The Fluorine Content of Solid FuelsSolid Fuels
EN 14582Characterization of Waste — Halogen and Sulfur Content — Oxygen Combustion in Closed Systems and Determination MethodsSolid, pasty and liquid waste samples

Specifications

Analytical Principle: High-temperature pyrolysis combined with ion chromatography

Sample Weight: ≤0.1 g

Measured Elements: Fluorine, chlorine, bromine, iodine and sulfur

Measurement Range:

  • Standard version: 5 ppm–1%
  • Ultra-low-level version: 0.2 ppm–1%

Pyrolysis Time: ≤20 min, adjustable

Pyrolysis Temperature: Up to 1250°C, adjustable

Temperature Control Accuracy: ±5°C

Maximum Power: 1.8 kW

Power Supply: AC 220 V ±10%, 50/60 Hz

Software & Schematic Diagram
On-site Photos
Videos

Our video library is currently under development, featuring product operation tutorials, application demonstrations, and technical explainers. Stay tuned for regular updates.

Customer Testimonials
We have provided a complete set of Sundy testing solutions, including bomb calorimeter, sulfur analyzer, and thermogravimetric analyzer, to the fuel analysis laboratory of Kavçim Cement, a leading integrated cement plant on Turkey’s Black Sea coast. Sundy’s standardized instrument performance bring outstanding practical value to end users. The upgraded solution significantly simplifies daily operations, lowers manual intervention and maintenance costs, and unifies data consistency for routine fuel analysis.We are very satisfied with Sundy’s product quality and technical support, and will continue to promote and recommend Sundy analytical instruments for more industrial laboratory projects in Turkey.
Dr. Karahan, POTANTEK
After more than 10 years of continuous cooperation and market verification, we confirm that Sundy’s products, especially its automatic calorimeters and elemental analyzers, have high precise temperature control, accurate testing results, and ultra-low failure rate. All products are fully compliant with international ISO and ASTM standard, completely meeting the testing demands of coal laboratories, power plants, industrial enterprises, universities and scientific research institutions.We highly recognize Hunan Sundy Science and Technology’s strong production capacity, standardized quality management system, professional technical support team and excellent business credibility. In all our cooperation cases, Sundy has always demonstrated professionalism, reliability and a customer-oriented service attitude.
Dr. Kim, YEONJIN
FAQs
The system is designed for the qualitative and quantitative determination of fluorine, chlorine, bromine, iodine and sulfur.
The SDXNS200H combines high-temperature pyrolysis with ion chromatography. The sample is progressively introduced into the heated furnace, where the target elements are converted into hydrogen halides and sulfur-containing gases. These products are absorbed into a collection solution and subsequently separated and quantified by ion chromatography.
Applicable samples include biomass, coal, petrochemicals, paper products, degradable materials, electronic and electrical products, soil, ores and cement. Both solid and liquid samples can be introduced.

The specified measuring ranges are:

  • Standard version: 5 ppm to 1%
  • Ultra-low-level version: 0.2 ppm to 1%

The practical range may also depend on sample mass, collection volume, dilution factor and the selected calibration range.

The general technical specification is ≤0.1 g. The operating procedure recommends approximately 0.05 ± 0.001 g for a typical air-dried solid sample.
The pyrolysis temperature is adjustable up to 1250°C, with a specified temperature-control accuracy of ±5°C. The furnace is designed for long-term operation at approximately 1200°C.
The manual identifies F, Cl, Br, I and S as measurable elements. However, the simultaneous-analysis capability is specifically stated for fluorine, chlorine, bromine and sulfur. The manual does not explicitly state that iodine is included in the same four-component simultaneous method.
The adjustable pyrolysis time is no more than 15 minutes. Under continuous testing conditions, the stated average test time is also no more than 15 minutes per sample.
The instrument supports selectable gas conditions involving argon, oxygen and steam-assisted mixed-gas operation. The atmosphere and flow parameters can be configured according to the sample type and analytical method.
Yes. Liquid samples can be introduced using the liquid autosampler or placed in a crucible. When using a crucible, the manual recommends adding quartz wool and dispensing the liquid onto the quartz wool to reduce direct contact with and corrosion of the crucible.
Ion chromatography separates the ions before quantification. This reduces mutual interference among components in the collection solution and generally provides lower detection limits, higher sensitivity, better repeatability and more reliable multi-component analysis.
  • Oxygen: ≥99.95%
  • Argon: ≥99.999%

For ultra-low-level analysis, gas purification traps should be installed to reduce blank contamination.

The pre-start procedure specifies approximately:

  • Oxygen: 400 mL/min
  • Carrier argon: 200 mL/min
  • Steam-assisted argon: 100 mL/min

These values should normally be established during installation and should not be changed arbitrarily.

The recommended ambient temperature is 20 ± 10°C, with a temperature variation of no more than 3°C during one day of testing and relative humidity no higher than 85%.

For measurements below 1 ppm, the room temperature should preferably remain below 22°C, with a variation of no more than 2°C. The laboratory should also be clean, ventilated and free from corrosive gases, dust, vibration, strong electromagnetic fields and significant air currents.

The installation requirements specify ultrapure water with:

  • Resistivity: ≥18 MΩ·cm
  • TOC: ≤50 μg/L
  • Chloride: ≤1 μg/L
  • Total silicon: ≤3 μg/L

Water quality is particularly important for low-level fluorine and chlorine measurements.

After startup, the software evaluates baseline noise and drift against preset limits. Baseline stabilization generally requires more than one hour. Calibration or sample analysis should begin only after both noise and drift meet the specified acceptance criteria.

The Blank Sample Experiment option must first be enabled in the experimental settings. A blank run is then created without adding a sample. The operator selects an acceptable blank result, and the corresponding peak areas are subtracted from subsequent sample results.

For trace-level testing, the combustion tube, sample box, crucible, collection system, water and gases must all be controlled carefully because they may contribute to the system blank.

The system supports several calibration approaches:

  1. Automatic standard-solution preparation when the automatic dilution function is installed.
  2. Direct solution analysis using individually prepared standards.
  3. Calibration through the normal sample-testing route, using an autosampler or standards introduced into crucibles.

After the standard chromatograms are acquired, component names and concentrations are assigned, the calibration points are imported, and the software fits either all component curves or an individual component curve.

The workstation provides:

  • Instrument preparation and automatic testing
  • Furnace, gas-flow and IC control
  • Standard-curve generation and backup
  • Chromatogram acquisition and peak processing
  • Blank correction and result recalculation
  • Relative standard deviation calculation
  • Historical-data searching
  • Excel export and report printing
  • User, role and permission management
  • Parameter and database backup
  • Fault, communication and operation logs
  • Consumable replacement reminders
  • LIMS connectivity

The operator should check:

  • Gas tightness of the combustion and collection system
  • Oxygen and argon supply and flow rates
  • Selection of the correct pyrolysis method
  • Condition of the combustion tube and collection bottle
  • Cleanliness of the flow path
  • IC pump pressure and flow stability
  • Chromatographic column condition
  • Suppressor operation
  • Conductivity-cell condition
  • Baseline noise and drift
  • Sample homogeneity, mass and dilution settings

Leaks, insufficient gas flow, an unsuitable method, contaminated water or reagents, air bubbles in the IC flow path and column deterioration are common causes of unstable results.

Related Products

SDCHN636 Carbon, Hydrogen and Nitrogen Analyzer

 SDBS200 Flash Point Tester

SDCXS100 Calorific Value and Halogen Analyzer

Have Questions? Contact Us Directly

Share your application details and our team will recommend a suitable configuration.