Determination of Moisture, Volatile Matter, Ash, and Fixed Carbon in Coal and Coke with Sundy SDTGA6000A Thermogravimetric Analyzer According to ASTM D7582

Published On: 2026-10-30
Total Views: 19

 

Introduction

Moisture, volatile matter, and ash are key parameters for assessing solid fuel quality. Moisture is used to calculate dry-basis results. Ash is used in the ultimate analysis calculation of oxygen by difference and for material balance and ash loading in boiler systems. Volatile matter indicates coke yield on carbonization, reflects combustion characteristics, and serves as a basis for purchasing and selling solid fuels.

Sundy SDTGA6000A determines moisture, volatile matter, ash, and fixed carbon in coal and coke in accordance with ASTM D7582. Following the standard’s sequential procedure, it analyzes 20 samples simultaneously in a single run, with fixed carbon calculated by difference.

Method Reference

ASTM D7582 Standard Test Methods for Proximate Analysis of Coal and Coke by Macro Thermogravimetric Analysis

Summary of Test Method

Moisture is determined by measuring the loss in mass of the analysis specimen of coal or coke when heated under specified conditions of temperature, time, atmosphere, specimen mass, and equipment specifications.

Volatile matter is determined by measuring the loss in mass of the analysis specimen of coal or coke when heated under rigidly controlled conditions. The measured mass loss is used to calculate the volatile matter after correcting for the moisture level.

Ash is determined by measuring the mass of the residue remaining after burning the coal or coke specimen under specified conditions of specimen mass, temperature, time, atmosphere, and equipment specifications.

Fixed carbon is a calculated value equal to the difference between 100 and the sum of the moisture, ash, and volatile matter, where all values are expressed on the same moisture reference base.

Sample Preparation

The analysis sample should be prepared in accordance with ASTM D2013/D2013M, which shall be the material pulverized to pass a 250 μm (No. 60 US Standard mesh) sieve.

Certified Reference Materials

Coal or coke material(s) meeting the requirements of analysis sample, with a certificate of analysis specifying the reference value and the uncertainty of the reference value. Reference materials used for the calibration of volatile matter shall include information on the certificate of analysis detailing the method(s) employed to determine volatile matter of the reference material.

Accessories

Ceramic Crucibles

Ceramic Crucible Lids (used for Volatile matter test)

Method Parameters

Moisture

ParameterValue
Method StepMoisture
Sample weight (g)1.0
Crucible LidsNo
Test Temperature (°C)107
Hold Time (min)60
AtmosphereAir/Oxygen/Nitrogen
Flow Rate (L/min)15
Final WeightAt Constancy
Constancy Window (min)9
Constancy Level (g)0.0005

Volatile Matter

ParameterValue
Method StepVolatile
Crucible LidsYes
Start Temperature (°C)107*
End Temperature (°C)950
Ramp Rate (°C/min)35
Hold Time (min)9
AtmosphereNitrogen
Flow Rate (L/min)15
Final WeightAt End of Step

 

*Note 1 – In the sequential determination of moisture, volatile matter and ash, the start temperature is 107 °C, which is ambient temperature in the individual determination of moisture.

Ash

ParameterValue
Method StepAsh
Crucible LidsNo
Start Temperature (°C)600*
End Temperature (°C)750/950*
Ramp Rate (°C/min)3/5*
Hold Time (min)120/180*
AtmosphereAir/Oxygen
Flow Rate (L/min)15
Final WeightAt Constancy
Constancy Window (min)9
Constancy Level (g)0.0005

 

*Note 2 – In the sequential determination of moisture, volatile matter and ash, start temperature is 600 °C, which is 107 °C in the sequential determination of moisture and ash, and ambient temperature in the individual determination of ash.

*Note 3 – For coal, end temperature is 750 °C, which is 950 °C for coke.

*Note 4 – Ramp rate should be changed with the different start temperature and end temperature, which should keep the total heat-up time within 1 h. In the sequential determination of Moisture and Ash and the individual determination of Ash, the heating programs vary, which is raising the temperature of the furnace at a rate such that it reaches 500 °C ±10 °C at the end of 1 h. For coal, continue heating the samples until the temperature rises from 500 °C to 750 °C ±15 °C at the end of 1 h. For coke, continue heating the samples until the temperature rises from 500 °C to 950 °C ±20 °C at the end of 1 h.

*Note 5 – For coal, hold time is 120 min, which is 180 min for coke.

Procedure

(1) Verify the instrument can meet all specifications in the method parameters with respect to gas type, gas flows, and heating rates prior to use. Prepare the software and crucible by referring to the instruction manual.

(2) The instrument shall be calibrated for the determination of volatile matter employing certified reference materials (using a minimum of three coal or coke reference materials). The calibration shall be performed at the same furnace ramp temperature as that used for the analysis.

(3) After calibration for volatile matter is complete, analyze samples following the procedure outlined in the instruction manual.

Typical Results

Representative results obtained with the method described above are reported in Tables 1-3.

Table 1 – Moisture (Drying gas – nitrogen)

Sample IDMoisture
%
Diff.
%
Avg.
%
Repeatability limit
%
AR19333.740.013.740.25
3.75
AR17311.380.011.380.25
1.39
AR17331.450.011.440.25
1.44
AR732*0.260.000.260.11
0.26

Table 2 – Ash (Dry basis)

Sample IDAsh
%
Diff.
%
Avg.
%
Repeatability limit
%
AR19337.41 7.420.017.420.19
AR173141.66 41.670.0141.66/
AR173317.52 17.530.0117.520.19
AR732*6.70 6.700.006.700.09

Table 3 – Volatile Matter (Dry basis)

Sample IDVolatile matter
%
Diff.
%
Avg.
%
Repeatability limit
%
AR193336.150.2136.260.36
36.36
AR173125.850.1225.910.36
25.97
AR17336.110.016.120.36
6.12
AR732*1.130.121.190.36
1.25

 

*Note 6 – AR732 is coke, while the remaining samples are coals.

Summary

Sundy SDTGA6000A provides a fully automated, high-throughput solution for the proximate analysis of coal and coke in compliance with ASTM D7582. In a single run, it determines moisture, volatile matter, and ash for multiple specimens simultaneously, with fixed carbon calculated by difference. Representative results are within the repeatability limits specified in ASTM D7582 Standard Test Methods for Proximate Analysis of Coal and Coke by Macro Thermogravimetric Analysis, confirming that the instrument delivers accurate results while significantly improving batch testing efficiency.